        高溫PH/DO電檢測報告

        更新時間:2016-11-28      點擊次數:1371

        KNICK 德工程師對于

        上海博取高溫PH/DO電檢測報告

         

        Short test: Sensors from China

         

        pH-Sensors

         

         

        CHP5806/K8S

        CHP5806/VP

        Sensor connection

        S8

        VP

        Serial number

        6050134

        6010021

        buffer value

        4,00

        Potential buffer 1 [mV]

        174

        174

        Temperature buffer 1 [°C]

        25 (manual)

        23,0

        Time buffer 1 [s]

        21

        31

        buffer value

        7,00

        Potential buffer 2 [mV]

        0

        0

        Temperature buffer 2 [°C]

        25 (manual)

        23,1

        Time buffer 2 [s]

        20

        25

         

        Zero Point [-]

        6,982

        7,000

        Slope [mV]

        58,6

        58,4

        Average Time [s]

        21

        29

        Imp. pH-electrode [MOhm]

        145,5

        Imp. Ref.-electrode [kOhm]

        17,5

         

        Functionality with Stratos Eco 2405PH: Yes.

                  Reference impedance above average.
                  CHP5806/K8S without integrated temperature sensor.

         

        Cond-Sensor DDG-0.1F

        Cell constant C = 0,121/cm
        (measured against a reference cell at 140 µS/cm and room temperature)
        Difference based on construction or polarization?
        Linearity and measurement range were not tested.

        Functionality with Stratos Eco 2405COND: Yes

        Oxy-Sensor DOG-208FA/KA12

        Sensor current: circa 80 nA (air, 25 °C, 1013 mbar)
        Zero point current: circa 0,065 nA (N2 5.0),  approximate resolution of 0,01 mg/l (10 µg/l)
        Temperature coefficient of membrane 3%/K?
        Drift behavior, reaction to sterilization, influence of pressure variation were not tested.
        Functionality with Stratos Eco 2405Oxy: Yes.
        Max. Resolution of Stratos Eco is 0,01 mg/l (ppm), no trace resolution!
        Sensor Typ A (default)
        Membrane compensation nedds to be set at 1,23 (default is 1,00)

        21. September 2016 sc

        高溫PH和DO電的檢測結果非常好。